Description
Product details
anti-human IgG is used to label ABEI, and use purified Rubella antigen to coat nano magnetic microbeads. Sample, Calibrator or Control with Buffer(goat Anti-human IgM, goat Anti-human IgA) and nano magnetic microbeads coated with Rubella antigen are mixed thoroughly and incubated at 37℃ and cycle washing for 1 time. Then add ABEI Label, incubation and form a sandwich, then washing for the 2nd time. Subsequently, the starter reagents are added and a flash chemiluminescent reaction is initiated. The light signal is measured by a photomultiplier as RLU within 3 seconds and is proportional to the concentration of Rubella IgG present in samples